Live Results TARA EUROPA-micro
To overview 2
EasyClus check & Info of last file

----- Processed at : 21-Jul-2023 16:46:01 -----

1. The cytometric analysis is performed on machine IFREMER
2. The CytoUSB software used is CytoUSB Version: 5.11.1.25
3. The filename is FLR26 2023-04-19 08h05.cyz
4. The analysis is performed on 19-04-2023 time 08:05:48
5. The number of measured particles is 95909
6. The total concentration found (e.g. only SWS trigg.) is 27 particles/microliter
7. The concentration found (all used triggers) is 27 particles/microliter
8. The preconcentration (before analysis) found (all used triggers) is 45 particles/microliter
9. The flush time found (between samples) is 60 seconds
10. The effective analysis volume is 3512.2826 microliter
11. The pumped volume is 5299.2 microliter
12. The particle-rate is 301 particles passing the laser beam/second
13. The used sample flow rate is 11.0405 microliter/second
14. The used block size memory is 4 kB
15. Trigger 1: FL Red at threshold level 26 mV
16. Trigger 2: not used at threshold level -999
17. Temperature sensor PMT is : 29.5 C
18. Temperature sensor Sheath is : 28.8 C
19. Temperature sensor Sheath is : 28.0 C
20. Temperature sensor System is : 26.0 C
21. Pressure sensor difference is : -271 mBar
22. Pressure sensor absolute is : 255 mBar
23. Sample pump 1: Last calibration of the sample pump was : 02-Févr-2023
24. Alignment check 1 by curvature detector is 0.72951 (1=perfect)
25. Alignment check 2 by curvature detector is 20% (100%=perfect)
26. Saturated signals (flattened peaks) occur when the dynamic range of the (HS)detector is reached
27. FWS: Nr of saturated signals is 0 (0% of 95909 measured particles)
28. FWS: Maximum signal height found is 2909 mV
29. FWS: Nr of particles with signal height <10 mV is high (174 part. (0 %))
30. SWS HS: Nr of saturated signals is 826 (0% of 95909 measured particles)
31. SWS HS: Maximum signal height found is 5000 mV
32. SWS HS: Nr of particles with signal height <10 mV is high (120 part. (0 %))
33. FL Red HS: Nr of saturated signals is 0 (0% of 95909 measured particles)
34. FL Red HS: Maximum signal height found is 4569 mV
35. FL Red HS: Nr of particles with signal height <10 mV is high (4 part. (0 %))
36. FL Yellow HS: Nr of saturated signals is 0 (0% of 95909 measured particles)
37. FL Yellow HS: Maximum signal height found is 70 mV
38. FL Yellow HS: Nr of particles with signal height <10 mV is high (95886 part. (100 %))
39. FL Orange HS: Nr of saturated signals is 0 (0% of 95909 measured particles)
40. FL Orange HS: Maximum signal height found is 2587 mV
41. FL Orange HS: Nr of particles with signal height <10 mV is high (82171 part. (86 %))
42. FL Green HS: Nr of saturated signals is 0 (0% of 95909 measured particles)
43. FL Green HS: Maximum signal height found is 958 mV
44. FL Green HS: Nr of particles with signal height <10 mV is high (95767 part. (100 %))
45. FWS: Nr of block shaped signals is 433 (0.45147%)
46. SWS HS: Nr of block shaped signals is 403 (0.42019%)
47. FL Red HS: Nr of block shaped signals is 18 (0.018768%)
48. FL Orange HS: Nr of block shaped signals is 11 (0.011469%)
49. FL Green HS: Nr of block shaped signals is 7 (0.0072986%)
50. Suspect nr: 0=low, 10=high
51. Maximum FWS: suspect = 0
52. Maximum SWS HS: suspect = 0
53. Maximum FL Red HS: suspect = 0
54. Maximum FL Yellow HS: suspect = 0
55. Maximum FL Orange HS: suspect = 2
56. Maximum FL Green HS: suspect = 2
57. SWS: PMT level is 60
58. FL Red: PMT level is 80
59. FL Yellow: PMT level is 90
60. FL Orange: PMT level is 94
61. FL Green: PMT level is 94

-------------------------- Warnings info --------------------------
1.03 Warning: Your preconcentration differs more than 25% from the during analysis concentration
1.03.1 Solution - Carry over between samples? > Flush longer
1.03.2 Solution - Better homogenize samples during subsampling by a gentle stirrer
1.03.3 Solution - Colonies-filaments of cells break during analysis > stir slower
1.05 Warning: Your effective analysis volume is low compared to the pumped analysis volume
1.05.1 Solution - Set blocksize memory in CytoUSB higher, used=4k (default=4k) e.g. to 8k
1.09 Warning: This (default) low block size of 4kB in CytoUSB will decrease the effective analysis volume. (Only older CytoSenses)
1.09.1 Solution - Increase block size memory if the total particle rate is high enough (but prevent time outs) OR set to 'auto'
1.11 Warning: This threshold level is quite high
1.11 Warning: Take care you don't miss smaller particles, because they will not be triggered at this level
1.11.1 Solution - Set lower threshold trigger level
1.11.2 Solution - If you need to put the level that high to measure something, it might be causeed by increased noise.
1.11.3 Solution - Increased noise is usually caused by dirty cuvette (inner) walls/tubing/sheath fluid - So, clean the cuvette carefully, change tubing, use new sheath
1.23 Warning: FWS-L and -R might indicate an air bubble problem 20% (100%=perfect)
1.23.1 Solution - Pinch valve to get rid of the air bubble or blockage by something different
1.26 Warning: FL Yellow HS: Biggest particles found is only 70 mV. Detector sensitivity might be too low.
1.26.1 Solution - Increase, set higher FL Yellow HS PMT voltage in CytoUSB (make more sensitive)
1.27 Warning: FL Yellow HS : Nr of particles with signal height <10 mV is high (100 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.27 Warning: FL Orange HS : Nr of particles with signal height <10 mV is high (86 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.27 Warning: FL Green HS : Nr of particles with signal height <10 mV is high (100 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
Disclaimer
Any use of the provided data and results or information provided by this website is on full own risk and responsibility. Those who gave access to this website or those who provided the results represented on the website cannot be held responsible for any damage or consequences that could arise from the use of this information. This LIVE system is empowered by the Rijkswaterstaat, CytoBuoy BV and Thomas Rutten Projects. TRP Copyright 2020