PhytoplanktonLIVE.com
Online automatic phytoplankton monitoring at several locations performed by flowcytometry (CytoSense-CytoSub)
< HOME  TO OVERVIEW  1 TO OVERVIEW  2 >

Live Results SCOPES
To overview 2
EasyClus check & Info of last file

----- Processed at : 14-Jan-2023 08:36:53 -----

1. The cytometric analysis is performed on machine LOG I
2. The CytoUSB software used is CytoUSB Version: 5.11.1.6
3. The filename is SCOPES_flr1_sws10_maxsws4_5ul_16min 2023-01-14 08h19.cyz
4. The analysis is performed on 14-01-2023 time 08:20:23
5. The number of measured particles is 76721
6. The total concentration found (e.g. only SWS trigg.) is 24 particles/microliter
7. The concentration found (all used triggers) is 14 particles/microliter
8. The preconcentration (before analysis) found (all used triggers) is 14 particles/microliter
9. The flush time found (between samples) is 60 seconds
10. The effective analysis volume is 5459.4922 microliter
11. The pumped volume is 5472 microliter
12. The particle-rate is 134 particles passing the laser beam/second
13. The used sample flow rate is 5.7015 microliter/second
14. The used block size memory is 4 kB
15. Trigger 1: FL Red at threshold level 1 mV
16. Trigger 2: Total SWS at threshold level 10
17. Trigger 3: Maximum SWS at threshold level 4
18. Temperature sensor PMT is : 25.8 C
19. Temperature sensor Sheath is : 24.5 C
20. Temperature sensor Sheath is : 23.9 C
21. Temperature sensor System is : 21.2 C
22. Pressure sensor difference is : -302 mBar
23. Pressure sensor absolute is : 209 mBar
24. Sample pump 1: Last calibration of the sample pump was : 19-Oct-2022
25. Alignment check 1 by curvature detector is 0.98551 (1=perfect)
26. Alignment check 2 by curvature detector is 31% (100%=perfect)
27. Saturated signals (flattened peaks) occur when the dynamic range of the (HS)detector is reached
28. FWS: Nr of saturated signals is 1 (0% of 76721 measured particles)
29. FWS: Maximum signal height found is 5120 mV
30. FWS: Nr of particles with signal height <10 mV is high (42243 part. (55 %))
31. SWS HS: Nr of saturated signals is 16 (0% of 76721 measured particles)
32. SWS HS: Maximum signal height found is 7360 mV
33. SWS HS: Nr of particles with signal height <10 mV is high (7436 part. (10 %))
34. FL Yellow HS: Nr of saturated signals is 0 (0% of 76721 measured particles)
35. FL Yellow HS: Maximum signal height found is 4000 mV
36. FL Yellow HS: Nr of particles with signal height <10 mV is high (58274 part. (76 %))
37. FL Orange HS: Nr of saturated signals is 0 (0% of 76721 measured particles)
38. FL Orange HS: Maximum signal height found is 1176 mV
39. FL Orange HS: Nr of particles with signal height <10 mV is high (63568 part. (83 %))
40. FL Red HS: Nr of saturated signals is 1 (0% of 76721 measured particles)
41. FL Red HS: Maximum signal height found is 5664 mV
42. FL Red HS: Nr of particles with signal height <10 mV is high (41755 part. (54 %))
43. FWS: Nr of block shaped signals is 17 (0.022158%)
44. SWS HS: Nr of block shaped signals is 16 (0.020855%)
45. FL Yellow HS: Nr of block shaped signals is 1 (0.0013034%)
46. Suspect nr: 0=low, 10=high
47. Maximum FWS: suspect = 0
48. Maximum SWS HS: suspect = 0
49. Maximum FL Yellow HS: suspect = 0
50. Maximum FL Orange HS: suspect = 0
51. Maximum FL Red HS: suspect = 0
52. SWS: PMT level is 60
53. FL Yellow: PMT level is 90
54. Fl Orange: PMT level is 85
55. FL Red: PMT level is 85

-------------------------- Warnings info --------------------------
1.09 Warning: This (default) low block size of 4kB in CytoUSB will decrease the effective analysis volume. (Only older CytoSenses)
1.09.1 Solution - Increase block size memory if the total particle rate is high enough (but prevent time outs) OR set to 'auto'
1.10 Warning: This threshold level is quite low and near noise
1.10 Warning: Take care you don't measure noise or in noise (this will give high particle rates but is in fact noise)
1.10.1 Solution - Increase threshold trigger level
1.23 Warning: FWS-L and -R might indicate an air bubble problem 31% (100%=perfect)
1.23.1 Solution - Pinch valve to get rid of the air bubble or blockage by something different
1.27 Warning: FWS : Nr of particles with signal height <10 mV is high (55 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.27 Warning: FL Yellow HS : Nr of particles with signal height <10 mV is high (76 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.27 Warning: FL Orange HS : Nr of particles with signal height <10 mV is high (83 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.27 Warning: FL Red HS : Nr of particles with signal height <10 mV is high (54 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
Disclaimer
Any use of the provided data and results or information provided by this website is on full own risk and responsibility. Those who gave access to this website or those who provided the results represented on the website cannot be held responsible for any damage or consequences that could arise from the use of this information. This LIVE system is empowered by the ULCO-CNRS, CytoBuoy BV and Thomas Rutten Projects. TRP Copyright 2023