Live Results RESILIENCE
To overview 2
EasyClus check & Info of last file

----- Processed at : 03-Jun-2022 19:49:08 -----

1. The cytometric analysis is performed on machine MIO - La Réunion
2. The CytoUSB software used is CytoUSB Version: 5.11.0.17
3. The filename is MAP-IO-Resilience_22-FLR5 2022-05-21 19h22.cyz
4. The analysis is performed on 21-05-2022 time 19:22:09
5. The number of measured particles is 16279
6. The total concentration found (e.g. only SWS trigg.) is 14 particles/microliter
7. The concentration found (all used triggers) is 14 particles/microliter
8. The preconcentration (before analysis) found (all used triggers) is 14 particles/microliter
9. The flush time found (between samples) is 100 seconds
10. The effective analysis volume is 1165.2256 microliter
11. The pumped volume is 1214.4 microliter
12. The particle-rate is 71 particles passing the laser beam/second
13. The used sample flow rate is 5.0606 microliter/second
14. The used block size memory is 4 kB
15. Trigger 1: FL Red at threshold level 5 mV
16. Trigger 2: not used at threshold level -999
17. Temperature sensor PMT is : 34.7 C
18. Temperature sensor Sheath is : 33.2 C
19. Temperature sensor Sheath is : 35.2 C
20. Temperature sensor System is : 31.3 C
21. Pressure sensor difference is : -232 mBar
22. Pressure sensor absolute is : 223 mBar
23. Sample pump 1: Last calibration of the sample pump was : 12-Oct-2020
24. Alignment check 1 by curvature detector is 1.1973 (1=perfect)
25. Alignment check 2 by curvature detector is 38% (100%=perfect)
26. Saturated signals (flattened peaks) occur when the dynamic range of the (HS)detector is reached
27. FWS: Nr of saturated signals is 3 (0% of 16279 measured particles)
28. FWS: Maximum signal height found is 13952 mV
29. FWS: Nr of particles with signal height <10 mV is high (13150 part. (81 %))
30. SWS HS: Nr of saturated signals is 3 (0% of 16279 measured particles)
31. SWS HS: Maximum signal height found is 7488 mV
32. SWS HS: Nr of particles with signal height <10 mV is high (12462 part. (77 %))
33. FL Yellow HS: Nr of saturated signals is 0 (0% of 16279 measured particles)
34. FL Yellow HS: Maximum signal height found is 4160 mV
35. FL Yellow HS: Nr of particles with signal height <10 mV is high (15133 part. (93 %))
36. FL Red HS: Nr of saturated signals is 1 (0% of 16279 measured particles)
37. FL Red HS: Maximum signal height found is 7488 mV
38. FL Red HS: Nr of particles with signal height <10 mV is high (13481 part. (83 %))
39. SWS HS: Nr of block shaped signals is 403 (2.4756%)
40. FWS: Nr of block shaped signals is 292 (1.7937%)
41. FL Yellow HS: Nr of block shaped signals is 10 (0.061429%)
42. Suspect nr: 0=low, 10=high
43. Maximum FWS: suspect = 0
44. Maximum SWS HS: suspect = 1
45. Maximum FL Yellow HS: suspect = 0
46. Maximum FL Red HS: suspect = 1
47. SWS: PMT level is 54
48. Fl Yellow: PMT level is 90
49. FL Red: PMT level is 95

-------------------------- Warnings info --------------------------
1.09 Warning: This (default) low block size of 4kB in CytoUSB will decrease the effective analysis volume
1.09.1 Solution - Increase block size memory if the total particle rate is high enough (but prevent time outs)
1.10 Warning: This threshold level is quite low and near noise
1.10 Warning: Take care you don't measure noise or in noise (this will give high particle rates but is in fact noise)
1.10.1 Solution - Increase threshold trigger level
1.21 Warning: Last calibration of sample pump towards analysis date was more than 90 days. (12-Oct-2020)
1.21.1 Solution - Check if the pumped analysis volume is the real volume by weighing
1.23 Warning: FWS-L and -R might indicate an air bubble problem 38% (100%=perfect)
1.23.1 Solution - Pinch valve to get rid of the air bubble or blockage by something different
1.27 Warning: FWS : Nr of particles with signal height <10 mV is high (81 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.27 Warning: SWS HS : Nr of particles with signal height <10 mV is high (77 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.27 Warning: FL Yellow HS : Nr of particles with signal height <10 mV is high (93 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.27 Warning: FL Red HS : Nr of particles with signal height <10 mV is high (83 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
Disclaimer
Any use of the provided data and results or information provided by this website is on full own risk and responsibility. Those who gave access to this website or those who provided the results represented on the website cannot be held responsible for any damage or consequences that could arise from the use of this information. This LIVE system is empowered by the ULCO-CNRS, CytoBuoy BV and Thomas Rutten Projects. TRP Copyright 2023