----- Processed at : 28-Jul-2023 12:25:06 -----

1. The cytometric analysis is performed on machine LOG I
2. The CytoUSB software used is CytoUSB Version: 5.11.1.6
3. The filename is APERO_Pico_continu_COURT_flr2_10ul_5min 2023-06-26 15h49.cyz
4. The analysis is performed on 26-06-2023 time 15:50:27
5. The number of measured particles is 10030
6. The total concentration found (e.g. only SWS trigg.) is 4 particles/microliter
7. The concentration found (all used triggers) is 4 particles/microliter
8. The preconcentration (before analysis) found (all used triggers) is 4 particles/microliter
9. The flush time found (between samples) is 60 seconds
10. The effective analysis volume is 2854.0082 microliter
11. The pumped volume is 2856 microliter
12. The particle-rate is 33 particles passing the laser beam/second
13. The used sample flow rate is 9.5194 microliter/second
14. The used block size memory is 4 kB
15. Trigger 1: FL Red at threshold level 2 mV
16. Trigger 2: not used at threshold level -999
17. Temperature sensor PMT is : 28.7 C
18. Temperature sensor Sheath is : 0.0 C
19. Temperature sensor Sheath is : 26.8 C
20. Temperature sensor System is : 24.0 C
21. Pressure sensor difference is : -400 mBar
22. Pressure sensor absolute is : 271 mBar
23. Sample pump 1: Last calibration of the sample pump was : 28-Févr-2023
24. Alignment check 1 by curvature detector is 0.8808 (1=perfect)
25. Alignment check 2 by curvature detector is 16% (100%=perfect)
26. Saturated signals (flattened peaks) occur when the dynamic range of the (HS)detector is reached
27. FWS: Nr of saturated signals is 1 (0% of 10030 measured particles)
28. FWS: Maximum signal height found is 11776 mV
29. FWS: Nr of particles with signal height <10 mV is high (936 part. (9 %))
30. SWS HS: Nr of saturated signals is 18 (0% of 10030 measured particles)
31. SWS HS: Maximum signal height found is 7488 mV
32. SWS HS: Nr of particles with signal height <10 mV is high (628 part. (6 %))
33. FL Yellow HS: Nr of saturated signals is 0 (0% of 10030 measured particles)
34. FL Yellow HS: Maximum signal height found is 824 mV
35. FL Yellow HS: Nr of particles with signal height <10 mV is high (9325 part. (93 %))
36. FL Orange HS: Nr of saturated signals is 0 (0% of 10030 measured particles)
37. FL Orange HS: Maximum signal height found is 146 mV
38. FL Orange HS: Nr of particles with signal height <10 mV is high (9804 part. (98 %))
39. FL Red HS: Nr of saturated signals is 1 (0% of 10030 measured particles)
40. FL Red HS: Maximum signal height found is 5920 mV
41. FL Red HS: Nr of particles with signal height <10 mV is high (1434 part. (14 %))
42. SWS HS: Nr of block shaped signals is 12 (0.11964%)
43. FWS: Nr of block shaped signals is 9 (0.089731%)
44. Suspect nr: 0=low, 10=high
45. Maximum FWS: suspect = 0
46. Maximum SWS HS: suspect = 1
47. Maximum FL Yellow HS: suspect = 3
48. Maximum FL Orange HS: suspect = 2
49. Maximum FL Red HS: suspect = 1
50. SWS: PMT level is 60
51. FL Yellow: PMT level is 90
52. Fl Orange: PMT level is 85
53. FL Red: PMT level is 85

-------------------------- Warnings info --------------------------
1.09 Warning: This (default) low block size of 4kB in CytoUSB will decrease the effective analysis volume. (Only older CytoSenses)
1.09.1 Solution - Increase block size memory if the total particle rate is high enough (but prevent time outs) OR set to 'auto'
1.10 Warning: This threshold level is quite low and near noise
1.10 Warning: Take care you don't measure noise or in noise (this will give high particle rates but is in fact noise)
1.10.1 Solution - Increase threshold trigger level
1.21 Warning: Last calibration of sample pump towards analysis date was more than 90 days. (28-Févr-2023)
1.21.1 Solution - Check if the pumped analysis volume is the real volume by weighing
1.23 Warning: FWS-L and -R might indicate an air bubble problem 16% (100%=perfect)
1.23.1 Solution - Pinch valve to get rid of the air bubble or blockage by something different
1.27 Warning: FL Yellow HS : Nr of particles with signal height <10 mV is high (93 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.26 Warning: FL Orange HS: Biggest particles found is only 146 mV. Detector sensitivity might be too low.
1.26.1 Solution - Increase, set higher FL Orange HS PMT voltage in CytoUSB (make more sensitive)
1.27 Warning: FL Orange HS : Nr of particles with signal height <10 mV is high (98 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.25 Warning: Suspected particles data blocks found!
1.29.1: Maximum FL Yellow HS: suspect = 3
1.29.11 Solution - Particle rate too high?
1.29.21 Solution - Trigger too sensitive?
1.29.31 Solution - Details look at 'DATA DIAGNOSIS' in 'Check CytoSense-Diagnosis-Meta Data'
Disclaimer
Any use of the provided data and results or information provided by this website is on full own risk and responsibility. Those who gave access to this website or those who provided the results represented on the website cannot be held responsible for any damage or consequences that could arise from the use of this information. This LIVE system is empowered by the Rijkswaterstaat, CytoBuoy BV and Thomas Rutten Projects. TRP Copyright 2020