PhytoplanktonLIVE.com
Online automatic phytoplankton monitoring at several locations performed by flowcytometry (CytoSense-CytoSub)
< HOME  TO OVERVIEW  1 TO OVERVIEW  2 >

Live Results AMAZOMIX
To overview 2
EasyClus check & Info of last file

----- Processed at : 30-Nov-2022 18:55:41 -----

1. The cytometric analysis is performed on machine LOG I
2. The CytoUSB software used is CytoUSB Version: 5.9.4.8
3. The filename is AMAZOMIX_Picophyto_clear_waters_continu 2021-09-22 00h40.cyz
4. The analysis is performed on 22-09-2021 time 00:43:20
5. The number of measured particles is 84285
6. The total concentration found (e.g. only SWS trigg.) is 1368 particles/microliter
7. The concentration found (all used triggers) is 41 particles/microliter
8. The preconcentration (before analysis) found (all used triggers) is 41 particles/microliter
9. The flush time found (between samples) is 70 seconds
10. The effective analysis volume is 2041.9735 microliter
11. The pumped volume is 2097.6 microliter
12. The particle-rate is 5977 particles passing the laser beam/second
13. The used sample flow rate is 4.3703 microliter/second
14. The used block size memory is 4 kB
15. Trigger 1: FL Red at threshold level 2 mV
16. Trigger 2: Length SWS at threshold level 1
17. Trigger 3: Maximum SWS at threshold level 3
18. Trigger 4: Maximum FWS at threshold level 2
19. Temperature sensor PMT is : 33.8 C
20. Temperature sensor Sheath is : 33.0 C
21. Temperature sensor Sheath is : 33.1 C
22. Temperature sensor System is : 29.9 C
23. Pressure sensor difference is : -244 mBar
24. Pressure sensor absolute is : 197 mBar
25. Sample pump 1: Last calibration of the sample pump was : 05-Mai-2021
26. Alignment check 1 by curvature detector is 0.82639 (1=perfect)
27. Alignment check 2 by curvature detector is 21% (100%=perfect)
28. Saturated signals (flattened peaks) occur when the dynamic range of the (HS)detector is reached
29. FWS: Nr of saturated signals is 454 (0% of 84285 measured particles)
30. FWS: Maximum signal height found is 14976 mV
31. FWS: Nr of particles with signal height <10 mV is high (53701 part. (64 %))
32. SWS HS: Nr of saturated signals is 704 (0% of 84285 measured particles)
33. SWS HS: Maximum signal height found is 7488 mV
34. SWS HS: Nr of particles with signal height <10 mV is high (2687 part. (3 %))
35. FL Yellow HS: Nr of saturated signals is 0 (0% of 84285 measured particles)
36. FL Yellow HS: Maximum signal height found is 4160 mV
37. FL Yellow HS: Nr of particles with signal height <10 mV is high (37509 part. (45 %))
38. FL Orange HS: Nr of saturated signals is 0 (0% of 84285 measured particles)
39. FL Orange HS: Maximum signal height found is 1296 mV
40. FL Orange HS: Nr of particles with signal height <10 mV is high (83423 part. (99 %))
41. FL Red HS: Nr of saturated signals is 5 (0% of 84285 measured particles)
42. FL Red HS: Maximum signal height found is 7488 mV
43. FL Red HS: Nr of particles with signal height <10 mV is high (63886 part. (76 %))
44. SWS HS: Nr of block shaped signals is 3333 (3.9544%)
45. FWS: Nr of block shaped signals is 2920 (3.4644%)
46. FL Yellow HS: Nr of block shaped signals is 242 (0.28712%)
47. Suspect nr: 0=low, 10=high
48. Maximum FWS: suspect = 2
49. Maximum SWS HS: suspect = 0
50. Maximum FL Yellow HS: suspect = 0
51. Maximum FL Orange HS: suspect = 1
52. Maximum FL Red HS: suspect = 0
53. SWS: PMT level is 65
54. FL Yellow: PMT level is 105
55. Fl Orange: PMT level is 95
56. FL Red: PMT level is 100

-------------------------- Warnings info --------------------------
1.07 Warning: Analysis with high particle rate (5977), meaning you have higher chance of coincidence
1.07.1 Solution - Decrease, use lower sample rate
1.07.2 Solution - Is threshold in noise level? Increase threshold and find out of particle rate is suddenly lower and quite stable
1.07.3 Solution - Is the sheath fluid clean enough?
1.09 Warning: This (default) low block size of 4kB in CytoUSB will decrease the effective analysis volume. (Only older CytoSenses)
1.09.1 Solution - Increase block size memory if the total particle rate is high enough (but prevent time outs) OR set to 'auto'
1.10 Warning: This threshold level is quite low and near noise
1.10 Warning: Take care you don't measure noise or in noise (this will give high particle rates but is in fact noise)
1.10.1 Solution - Increase threshold trigger level
1.21 Warning: Last calibration of sample pump towards analysis date was more than 90 days. (05-Mai-2021)
1.21.1 Solution - Check if the pumped analysis volume is the real volume by weighing
1.23 Warning: FWS-L and -R might indicate an air bubble problem 21% (100%=perfect)
1.23.1 Solution - Pinch valve to get rid of the air bubble or blockage by something different
1.27 Warning: FWS : Nr of particles with signal height <10 mV is high (64 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.27 Warning: FL Orange HS : Nr of particles with signal height <10 mV is high (99 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.27 Warning: FL Red HS : Nr of particles with signal height <10 mV is high (76 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
Disclaimer
Any use of the provided data and results or information provided by this website is on full own risk and responsibility. Those who gave access to this website or those who provided the results represented on the website cannot be held responsible for any damage or consequences that could arise from the use of this information. This LIVE system is empowered by the ULCO-CNRS, CytoBuoy BV and Thomas Rutten Projects. TRP Copyright 2023