Live Results AMATLANTE-C
To overview 2
EasyClus check & Info of last file

----- Processed at : 06-Aug-2021 22:57:43 -----

1. The cytometric analysis is performed on machine LOG I
2. The CytoUSB software used is CytoUSB Version: 5.9.4.8
3. The filename is AMATLANTE_flr2_5ul_8min_reajusted 2021-07-15 19h19.cyz
4. The analysis is performed on 15-07-2021 time 19:20:56
5. The number of measured particles is 24687
6. The total concentration found (e.g. only SWS trigg.) is 31 particles/microliter
7. The concentration found (all used triggers) is 10 particles/microliter
8. The preconcentration (before analysis) found (all used triggers) is 10 particles/microliter
9. The flush time found (between samples) is 70 seconds
10. The effective analysis volume is 2468.9092 microliter
11. The pumped volume is 2486.4 microliter
12. The particle-rate is 161 particles passing the laser beam/second
13. The used sample flow rate is 5.1821 microliter/second
14. The used block size memory is 4 kB
15. Trigger 1: FL Red at threshold level 2 mV
16. Trigger 2: Length SWS at threshold level 1
17. Trigger 3: Maximum SWS at threshold level 3
18. Trigger 4: Maximum FWS at threshold level 2
19. Temperature sensor PMT is : 30.9 C
20. Temperature sensor Sheath is : 29.7 C
21. Temperature sensor Sheath is : 28.5 C
22. Temperature sensor System is : 25.2 C
23. Pressure sensor difference is : -252 mBar
24. Pressure sensor absolute is : 191 mBar
25. Sample pump 1: Last calibration of the sample pump was : 05-May-2021
26. Alignment check 1 by curvature detector is 0.90146 (1=perfect)
27. Alignment check 2 by curvature detector is 66% (100%=perfect)
28. Saturated signals (flattened peaks) occur when the dynamic range of the (HS)detector is reached
29. FWS: Nr of saturated signals is 263 (1% of 24687 measured particles)
30. FWS: Maximum signal height found is 14976 mV
31. FWS: Nr of particles with signal height <10 mV is high (6249 part. (25 %))
32. SWS HS: Nr of saturated signals is 281 (1% of 24687 measured particles)
33. SWS HS: Maximum signal height found is 7488 mV
34. SWS HS: Nr of particles with signal height <10 mV is high (1102 part. (4 %))
35. FL Yellow HS: Nr of saturated signals is 0 (0% of 24687 measured particles)
36. FL Yellow HS: Maximum signal height found is 2624 mV
37. FL Yellow HS: Nr of particles with signal height <10 mV is high (14389 part. (58 %))
38. FL Orange HS: Nr of saturated signals is 0 (0% of 24687 measured particles)
39. FL Orange HS: Maximum signal height found is 43 mV
40. FL Orange HS: Nr of particles with signal height <10 mV is high (24674 part. (100 %))
41. FL Red HS: Nr of saturated signals is 8 (0% of 24687 measured particles)
42. FL Red HS: Maximum signal height found is 7488 mV
43. FL Red HS: Nr of particles with signal height <10 mV is high (17805 part. (72 %))
44. SWS HS: Nr of block shaped signals is 1056 (4.2776%)
45. FWS: Nr of block shaped signals is 929 (3.7631%)
46. FL Yellow HS: Nr of block shaped signals is 97 (0.39292%)
47. Suspect nr: 0=low, 10=high
48. Maximum FWS: suspect = 2
49. Maximum SWS HS: suspect = 0
50. Maximum FL Yellow HS: suspect = 2
51. Maximum FL Orange HS: suspect = 0
52. Maximum FL Red HS: suspect = 0
53. SWS: PMT level is 65
54. FL Yellow: PMT level is 105
55. Fl Orange: PMT level is 95
56. FL Red: PMT level is 100

-------------------------- Warnings info --------------------------
1.09 Warning: This (default) low block size of 4kB in CytoUSB will decrease the effective analysis volume
1.09.1 Solution - Increase block size memory if the total particle rate is high enough (but prevent time outs)
1.10 Warning: This threshold level is quite low and near noise
1.10 Warning: Take care you don't measure noise or in noise (this will give high particle rates but is in fact noise)
1.10.1 Solution - Increase threshold trigger level
1.23 Warning: FWS-L and -R might indicate an air bubble problem 66% (100%=perfect)
1.23.1 Solution - Pinch valve to get rid of the air bubble or blockage by something different
1.27 Warning: FL Yellow HS : Nr of particles with signal height <10 mV is high (58 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.26 Warning: FL Orange HS: Biggest particles found is only 43 mV. Detector sensitivity might be too low.
1.26.1 Solution - Increase, set higher FL Orange HS PMT voltage in CytoUSB (make more sensitive)
1.27 Warning: FL Orange HS : Nr of particles with signal height <10 mV is high (100 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.27 Warning: FL Red HS : Nr of particles with signal height <10 mV is high (72 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
Disclaimer
Any use of the provided data and results or information provided by this website is on full own risk and responsibility. Those who gave access to this website or those who provided the results represented on the website cannot be held responsible for any damage or consequences that could arise from the use of this information. This LIVE system is empowered by the ULCO-CNRS, CytoBuoy BV and Thomas Rutten Projects. TRP Copyright 2023