Fytoplankton.nl
Online automatic phytoplankton monitoring at several locations performed by flowcytometry (CytoSense-CytoSub)
< HOME  TO OVERVIEW  1 TO OVERVIEW  2 >

Live Results Endeavour
To overview 2
LL_Cend12_16_sws15flr6_3u_lIMG_600s 2016-06-26 12u21.cyz

----- Processed at : 30-Jun-2016 10:38:50 -----

1. The cytometric analysis is performed on machine LelystadII
2. The CytoUSB software used is CytoUSB Version: 5.7.5.5
3. The filename is LL_Cend12_16_sws15flr6_3u_lIMG_600s 2016-06-26 12u21.cyz
4. The analysis is performed on 26-06-2016 time 12:21:51
5. The number of measured particles is 13346
6. The total concentration found (e.g. only SWS trigg.) is 104 particles/microliter
7. The concentration found (all used triggers) is 9 particles/microliter
8. The preconcentration (before analysis) found (all used triggers) is 11 particles/microliter
9. The flush time found (between samples) is 120 seconds
10. The effective analysis volume is 1556.9588 microliter
11. The pumped volume is 1812 microliter
12. The particle-rate is 313 particles passing the laser beam/second
13. The used sample flow rate is 3.0184 microliter/second
14. The used block size memory is 64 kB
15. Trigger 1: SWS HS at threshold level 15 mV
16. Trigger 2: Maximum FL Red HS at threshold level 6 mV
17. Temperature sensor PMT is : 32 C
18. Temperature sensor Sheath is : 26 C
19. Temperature sensor Sheath is : 32 C
20. Pressure sensor difference is : -272 mBar
21. Pressure sensor absolute is : 180 mBar
22. Sample pump 1: Last calibration of the sample pump was : 18-Apr-2016
23. Alignment check 1 by curvature detector is 0.89551 (1=perfect)
24. Alignment check 2 by curvature detector is 80% (100%=perfect)
25. Saturated signals (flattened peaks) occur when the dynamic range of the (HS)detector is reached
26. FWS: Nr of saturated signals is 1 (0% of 13346 measured particles)
27. FWS: Maximum signal height found is 5528 mV
28. FWS: Nr of particles with signal height <10 mV is high (127 part. (1 %))
29. SWS HS: Nr of saturated signals is 18 (0% of 13346 measured particles)
30. SWS HS: Maximum signal height found is 5000 mV
31. SWS HS: Nr of particles with signal height <10 mV is high (0 part. (0 %))
32. FL Yellow HS: Nr of saturated signals is 15 (0% of 13346 measured particles)
33. FL Yellow HS: Maximum signal height found is 5000 mV
34. FL Yellow HS: Nr of particles with signal height <10 mV is high (12551 part. (94 %))
35. FL Orange HS: Nr of saturated signals is 15 (0% of 13346 measured particles)
36. FL Orange HS: Maximum signal height found is 5000 mV
37. FL Orange HS: Nr of particles with signal height <10 mV is high (12397 part. (93 %))
38. FL Red HS: Nr of saturated signals is 9 (0% of 13346 measured particles)
39. FL Red HS: Maximum signal height found is 5000 mV
40. FL Red HS: Nr of particles with signal height <10 mV is high (1418 part. (11 %))
41. FL Yellow HS: Nr of block shaped signals is 31 (0.23228%)
42. SWS HS: Nr of block shaped signals is 18 (0.13487%)
43. FWS: Nr of block shaped signals is 1 (0.0074929%)
44. FL Orange HS: Nr of block shaped signals is 1 (0.0074929%)
45. SWS HS: PMT level is 50
46. SWS LS: PMT level is 32
47. FL Yellow LS: PMT level is 61
48. FL Red LS: PMT level is 63
49. FL Yellow HS: PMT level is 83
50. FL Orange HS: PMT level is 80
51. FL Red HS: PMT level is 93
52. IIF: Imaging in Flow pictures available. Nr of images is 157

-------------------------- Warnings info --------------------------
1.20 Warning: FL Yellow LS: HS-LS ratio is 79 and too low.
1.20.1 Solution - Put LS PMT lower. Ratio HS/LS should be around 100.
1.20 Warning: FL Red LS: Maximum is only 20 mV (if HS is around 4500 mV) and too low.
1.20.1 Solution - Put LS PMT higher. Maximum LS around 50 mV (if HS is around 4500 mV).
1.20 Warning: FL Red LS: HS-LS ratio is 191 and too high.
1.20.1 Solution - Put LS PMT higher. Ratio HS/LS should be around 100.
1.23 Warning: FL Yellow HS : Nr of particles with signal height <10 mV is high (94 %)
1.23.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.23.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.23.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.23 Warning: FL Orange HS : Nr of particles with signal height <10 mV is high (93 %)
1.23.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.23.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.23.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
Disclaimer
Any use of the provided data and results or information provided by this website is on full own risk and responsibility. Those who gave access to this website or those who provided the results represented on the website cannot be held responsible for any damage or consequences that could arise from the use of this information. Powered by TRP © 2018