EIJ_SWS15_FLRed6_auto_uls_phyto_600s 2018-11-23 14h02.cyz

----- Processed at : 23-Nov-2018 14:13:04 -----

1. The cytometric analysis is performed on machine Eijsden
2. The CytoUSB software used is CytoUSB Version: 5.7.9.20
3. The filename is EIJ_SWS15_FLRed6_auto_uls_phyto_600s 2018-11-23 14h02.cyz
4. The analysis is performed on 23-11-2018 time 14:02:49
5. The number of measured particles is 566
6. The total concentration found (e.g. only SWS trigg.) is 70 particles/microliter
7. The concentration found (all used triggers) is 1 particles/microliter
8. The preconcentration (before analysis) found (all used triggers) is 0 particles/microliter
9. The flush time found (between samples) is 130 seconds
10. The effective analysis volume is 1037.8424 microliter
11. The pumped volume is 2748 microliter
12. The particle-rate is 320 particles passing the laser beam/second
13. The used sample flow rate is 4.5842 microliter/second
14. The used block size memory is 256 kB
15. Trigger 1: SWS HS at threshold level 15 mV
16. Trigger 2: Maximum FL Red HS at threshold level 6 mV
17. Temperature sensor PMT is : 30 C
18. Temperature sensor Sheath is : 26 C
19. Temperature sensor Sheath is : 30 C
20. Pressure sensor difference is : -207 mBar
21. Pressure sensor absolute is : 247 mBar
22. Sample pump 1: Last calibration of the sample pump was : 02-Aug-2018
23. Alignment check 1 by curvature detector is 1.0608 (1=perfect)
24. Alignment check 2 by curvature detector is 59% (100%=perfect)
25. Saturated signals (flattened peaks) occur when the dynamic range of the (HS)detector is reached
26. FWS: Nr of saturated signals is 1 (0% of 566 measured particles)
27. FWS: Maximum signal height found is 9911 mV
28. FWS: Nr of particles with signal height <10 mV is high (29 part. (5 %))
29. SWS HS: Nr of saturated signals is 0 (0% of 566 measured particles)
30. SWS HS: Maximum signal height found is 3745 mV
31. SWS HS: Nr of particles with signal height <10 mV is high (2 part. (0 %))
32. FL Yellow HS: Nr of saturated signals is 0 (0% of 566 measured particles)
33. FL Yellow HS: Maximum signal height found is 57 mV
34. FL Yellow HS: Nr of particles with signal height <10 mV is high (491 part. (87 %))
35. FL Orange HS: Nr of saturated signals is 0 (0% of 566 measured particles)
36. FL Orange HS: Maximum signal height found is 72 mV
37. FL Orange HS: Nr of particles with signal height <10 mV is high (516 part. (91 %))
38. FL Red HS: Nr of saturated signals is 0 (0% of 566 measured particles)
39. FL Red HS: Maximum signal height found is 124 mV
40. FL Red HS: Nr of particles with signal height <10 mV is high (448 part. (79 %))
41. FWS: Nr of block shaped signals is 151 (26.6784%)
42. SWS HS: Nr of block shaped signals is 52 (9.1873%)
43. FL Yellow HS: Nr of block shaped signals is 23 (4.0636%)
44. FL Orange HS: Nr of block shaped signals is 9 (1.5901%)
45. SWS LS: Nr of block shaped signals is 1 (0.17668%)
46. Suspect nr: 0=low, 10=high
47. Maximum FWS: suspect = 4
48. Maximum SWS HS: suspect = 4
49. Maximum FL Yellow HS: suspect = 2
50. Maximum FL Orange HS: suspect = 3
51. Maximum FL Red HS: suspect = 4
52. Maximum SWS LS: suspect = 0
53. Maximum FL Yellow LS: suspect = 2
54. Maximum FL Red LS: suspect = 0
55. SWS HS: PMT level is 56
56. FL Yellow HS: PMT level is 79
57. FL Orange HS: PMT level is 78
58. FL Red HS: PMT level is 97
59. SWS LS: PMT level is 40
60. FL Yellow LS: PMT level is 64
61. FL Red LS: PMT level is 73

-------------------------- Warnings info --------------------------
1.01 Warning: The number of saved particles in this analysis is quite low and
1.01 Warning: you might miss (statistically) bigger particles in your sample
1.01.1 Solution - Increase number of measured particles. Increase setted time OR after setted number of analyzed particles (in CytoUSB)
1.03 Warning: Your preconcentration differs more than 25% from the during analysis concentration
1.03.1 Solution - Carry over between samples? > Flush longer
1.03.2 Solution - Better homogenize samples during subsampling by a gentle stirrer
1.03.3 Solution - Colonies-filaments of cells break during analysis > stir slower
1.05 Warning: Your effective analysis volume is low compared to the pumped analysis volume
1.05.1 Solution - Set blocksize memory in CytoUSB higher, used=256k (default=4k) e.g. to 512k
1.17 Warning: Last calibration of sample pump towards analysis date was more than 90 days. (02-Aug-2018)
1.17.1 Solution - Check if the pumped analysis volume is the real volume by weighing
1.19 Warning: FWS-L and -R might indicate an air bubble problem 59% (100%=perfect)
1.19.1 Solution - Pinch valve to get rid of the air bubble or blockage by something different
1.20 Warning: SWS LS: HS-LS ratio is 29 and too low.
1.20.1 Solution - Put LS PMT lower. Ratio HS/LS should be around 100.
1.20 Warning: FL Yellow LS: HS-LS ratio is 999 and too high.
1.20.1 Solution - Put LS PMT higher. Ratio HS/LS should be around 100.
1.20 Warning: FL Yellow LS: HS-LS ratio is 999 and too high.
1.20.1 Solution - Put LS PMT higher. Ratio HS/LS should be around 100.
1.20 Warning: FL Red LS: HS-LS ratio is 999 and too high.
1.20.1 Solution - Put LS PMT higher. Ratio HS/LS should be around 100.
1.22 Warning: FL Yellow HS: Biggest particles found is only 57 mV. Detector sensitivity might be too low.
1.22.1 Solution - Increase, set higher FL Yellow HS PMT voltage in CytoUSB (make more sensitive)
1.23 Warning: FL Yellow HS : Nr of particles with signal height <10 mV is high (87 %)
1.23.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.23.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.23.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.22 Warning: FL Orange HS: Biggest particles found is only 72 mV. Detector sensitivity might be too low.
1.22.1 Solution - Increase, set higher FL Orange HS PMT voltage in CytoUSB (make more sensitive)
1.23 Warning: FL Orange HS : Nr of particles with signal height <10 mV is high (91 %)
1.23.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.23.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.23.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.22 Warning: FL Red HS: Biggest particles found is only 124 mV. Detector sensitivity might be too low.
1.22.1 Solution - Increase, set higher FL Red HS PMT voltage in CytoUSB (make more sensitive)
1.23 Warning: FL Red HS : Nr of particles with signal height <10 mV is high (79 %)
1.23.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.23.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.23.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.24 Warning: FWS: A relative large amount of particles are (rather) block shaped (27%)
1.24.1 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.24.2 Solution - Decrease, set lower FWS PMT voltage in CytoUSB (make less sensitive)
1.24 Warning: SWS HS: A relative large amount of particles are (rather) block shaped (9%)
1.24.1 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.24.2 Solution - Decrease, set lower SWS HS PMT voltage in CytoUSB (make less sensitive)
1.25 Warning: Suspected particles data blocks found!
1.25.1: Maximum FWS: suspect = 4
1.25.1 Solution - Particle rate too high?
1.25.1 Solution - Trigger too sensitive?
1.25.1 Solution - Details look at 'DATA DIAGNOSIS' in 'Check CytoSense-Diagnosis-Meta Data'
1.25.2: Maximum SWS HS: suspect = 4
1.25.2 Solution - Particle rate too high?
1.25.2 Solution - Trigger too sensitive?
1.25.2 Solution - Details look at 'DATA DIAGNOSIS' in 'Check CytoSense-Diagnosis-Meta Data'
1.25.3: Maximum FL Orange HS: suspect = 3
1.25.3 Solution - Particle rate too high?
1.25.3 Solution - Trigger too sensitive?
1.25.3 Solution - Details look at 'DATA DIAGNOSIS' in 'Check CytoSense-Diagnosis-Meta Data'
1.25.4: Maximum FL Red HS: suspect = 4
1.25.4 Solution - Particle rate too high?
1.25.4 Solution - Trigger too sensitive?
1.25.4 Solution - Details look at 'DATA DIAGNOSIS' in 'Check CytoSense-Diagnosis-Meta Data'
Disclaimer
Any use of the provided data and results or information provided by this website is on full own risk and responsibility. Those who gave access to this website or those who provided the results represented on the website cannot be held responsible for any damage or consequences that could arise from the use of this information. Powered by TRP © 2016