Online automatic phytoplankton monitoring at several locations performed by flowcytometry (CytoSense-CytoSub)
<
HOME
TO OVERVIEW 1
TO OVERVIEW 2
>
Live Results livelocation
To overview 2
EasyClus check & Info of last file
----- Processed at : 01-Nov-2024 17:11:43 -----
1. The cytometric analysis is performed on machine Cefas III
2. The CytoBuoy instrument serial-nr is CS-2022-114
3. The CytoUSB software used is CytoUSB Version: Build: 'ReleaseWebsite' , Commit: CytoUsb-v6.3.3.0-0-g9b2786abc, User=rob@Tommie, Machine=TOMMIE, ProjectPath=C:\Users\rob\Documents\Git\CytoSoftware\CytoUSB\+9b2786abc537a909b242db2a87f0202103b704b1
4. LASER : OBIS LS 488-60mW : Used at 50 mW
5. The filename is nano_cend17_24 2024-11-01 15h59.cyz
6. The analysis is performed on 01-11-2024 time 16:03:09
7. The number of measured particles is 133588
8. The total concentration found (e.g. only SWS trigg.) is 25 particles/microliter
9. The concentration found (all used triggers) is 25 particles/microliter
10. The preconcentration (before analysis) found (all used triggers) is 25 particles/microliter
11. The flush time found (between samples) is 190 seconds
12. The effective analysis volume is 5374.8885 microliter
13. The pumped volume is 5394 microliter
14. The particle-rate is 223 particles passing the laser beam/second
15. The used sample flow rate is 8.9854 microliter/second
16. Trigger 1: Fl Red at threshold level 3 mV
17. Trigger 2: not used at threshold level -999
18. Temperature sensor PMT is : 27.5 C
19. Temperature sensor Sheath is : 25.5 C
20. Temperature sensor Laser is : 26.6 C
21. Temperature sensor System is : 24.3 C
22. Pressure sensor difference is : -236 mBar
23. Pressure sensor absolute is : 281 mBar
24. Sample pump 1: Last calibration of the sample pump was : 05-Sept-2024
25. Alignment check 1 by curvature detector is 1.1158 (1=perfect)
26. Alignment check 2 by curvature detector is 73.2% (>70%=perfect)
27. Saturated signals (flattened peaks) occur when the dynamic range of the (HS)detector is reached
28. FWS: Nr of saturated signals is 1 (0% of 133588 measured particles)
29. FWS: Maximum signal height found is 11328 mV
30. FWS: Nr of particles with signal height <10 mV is high (16352 part. (12 %))
31. SWS HS: Nr of saturated signals is 1791 (0% of 133588 measured particles)
32. SWS HS: Maximum signal height found is 7488 mV
33. SWS HS: Nr of particles with signal height <10 mV is high (9395 part. (7 %))
34. FL Yellow HS: Nr of saturated signals is 0 (0% of 133588 measured particles)
35. FL Yellow HS: Maximum signal height found is 1280 mV
36. FL Yellow HS: Nr of particles with signal height <10 mV is high (101353 part. (76 %))
37. FL Orange HS: Nr of saturated signals is 0 (0% of 133588 measured particles)
38. FL Orange HS: Maximum signal height found is 1168 mV
39. FL Orange HS: Nr of particles with signal height <10 mV is high (129203 part. (97 %))
40. FL Red HS: Nr of saturated signals is 1 (0% of 133588 measured particles)
41. FL Red HS: Maximum signal height found is 7488 mV
42. FL Red HS: Nr of particles with signal height <10 mV is high (69308 part. (52 %))
43. FWS: Nr of block shaped signals is 3053 (2.2854%)
44. SWS HS: Nr of block shaped signals is 2619 (1.9605%)
45. Suspect nr: 0=low, 10=high
46. Maximum FWS: suspect = 0
47. Maximum SWS HS: suspect = 0
48. Maximum FL Yellow HS: suspect = 0
49. Maximum FL Orange HS: suspect = 0
50. Maximum FL Red HS: suspect = 0
51. Sidewards Scatter: PMT level is 60
52. Fl Yellow: PMT level is 95
53. Fl Orange: PMT level is 85
54. Fl Red: PMT level is 85
55. IIF: Imaging in Flow pictures available. Nr of images is 250
-------------------------- Warnings info --------------------------
1.10 Warning: This threshold level is quite low and near noise
1.10 Warning: Take care you don't measure noise or in noise (this will give high particle rates but is in fact noise)
1.10.1 Solution - Increase threshold trigger level
1.27 Warning: FL Yellow HS : Nr of particles with signal height <10 mV is high (76 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.27 Warning: FL Orange HS : Nr of particles with signal height <10 mV is high (97 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.27 Warning: FL Red HS : Nr of particles with signal height <10 mV is high (52 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)