PhytoplanktonLIVE.com
Online automatic phytoplankton monitoring at several locations performed by flowcytometry (CytoSense-CytoSub)
< HOME  TO OVERVIEW  1 TO OVERVIEW  2 >

To overview 2
EasyClus check & Info of last file

----- Processed at : 30-Oct-2023 09:17:14 -----

1. The cytometric analysis is performed on machine Cefas
2. The CytoUSB software used is CytoUSB Version: 5.7.1.0
3. The filename is nano_cend15_23 2023-10-30 09u06.cyz
4. The analysis is performed on 30-10-2023 time 09:06:59
5. The number of measured particles is 60027
6. The total concentration found (e.g. only SWS trigg.) is 18 particles/microliter
7. The concentration found (all used triggers) is 18 particles/microliter
8. The preconcentration (before analysis) found (all used triggers) is -1 particles/microliter
9. The flush time found (between samples) is 180 seconds
10. The effective analysis volume is 3411.3023 microliter
11. The pumped volume is 5160 microliter
12. The particle-rate is 151 particles passing the laser beam/second
13. The used sample flow rate is 8.6024 microliter/second
14. The used block size memory is 4 kB
15. Trigger 1: FL Red at threshold level 19 mV
16. Trigger 2: not used at threshold level -999
17. Temperature sensor PMT is : -999.0 C
18. Temperature sensor Sheath is : -999.0 C
19. Temperature sensor Sheath is : 24.3 C
20. Temperature sensor System is : -999.0 C
21. Pressure sensor difference is : -999 mBar
22. Pressure sensor absolute is : -999 mBar
23. Sample pump 1: Last calibration of the sample pump was : 01-Jan-0001
24. Alignment check 1 by curvature detector is 1.265 (1=perfect)
25. Alignment check 2 by curvature detector is 38% (100%=perfect)
26. Saturated signals (flattened peaks) occur when the dynamic range of the (HS)detector is reached
27. FWS: Nr of saturated signals is 0 (0% of 60027 measured particles)
28. FWS: Maximum signal height found is 4691 mV
29. FWS: Nr of particles with signal height <10 mV is high (15598 part. (26 %))
30. SWS HS: Nr of saturated signals is 0 (0% of 60027 measured particles)
31. SWS HS: Maximum signal height found is 4289 mV
32. SWS HS: Nr of particles with signal height <10 mV is high (55010 part. (92 %))
33. FL Yellow HS: Nr of saturated signals is 0 (0% of 60027 measured particles)
34. FL Yellow HS: Maximum signal height found is 837 mV
35. FL Yellow HS: Nr of particles with signal height <10 mV is high (53914 part. (90 %))
36. FL Orange HS: Nr of saturated signals is 0 (0% of 60027 measured particles)
37. FL Orange HS: Maximum signal height found is 2756 mV
38. FL Orange HS: Nr of particles with signal height <10 mV is high (45285 part. (75 %))
39. FL Red HS: Nr of saturated signals is 0 (0% of 60027 measured particles)
40. FL Red HS: Maximum signal height found is 4289 mV
41. FL Red HS: Nr of particles with signal height <10 mV is high (82 part. (0 %))
42. FWS: Nr of block shaped signals is 43 (0.071634%)
43. FL Yellow HS: Nr of block shaped signals is 29 (0.048312%)
44. SWS HS: Nr of block shaped signals is 28 (0.046646%)
45. FL Red HS: Nr of block shaped signals is 7 (0.011661%)
46. Suspect nr: 0=low, 10=high
47. Maximum FWS: suspect = 0
48. Maximum SWS HS: suspect = 0
49. Maximum FL Yellow HS: suspect = 0
50. Maximum FL Orange HS: suspect = 0
51. Maximum FL Red HS: suspect = 0
52. SWS HS: PMT level is 0
53. FL Yellow HS: PMT level is 0
54. FL Orange HS: PMT level is 0
55. FL Red HS: PMT level is 0

-------------------------- Warnings info --------------------------
1.03 Warning: Your preconcentration differs more than 25% from the during analysis concentration
1.03.1 Solution - Carry over between samples? > Flush longer
1.03.2 Solution - Better homogenize samples during subsampling by a gentle stirrer
1.03.3 Solution - Colonies-filaments of cells break during analysis > stir slower
1.05 Warning: Your effective analysis volume is low compared to the pumped analysis volume
1.05.1 Solution - Set blocksize memory in CytoUSB higher, used=4k (default=4k) e.g. to 8k
1.09 Warning: This (default) low block size of 4kB in CytoUSB will decrease the effective analysis volume. (Only older CytoSenses)
1.09.1 Solution - Increase block size memory if the total particle rate is high enough (but prevent time outs) OR set to 'auto'
1.16 Warning: Pressure difference is high.
1.16.1 Solution - Check if there is something blocking from sample inlet to sheath (filter block, tubing)
1.17 Warning: Inner absolute pressure in fluidics towards outside is high.
1.17.1 Solution - Check if there is something blocking in the fluidics pathways (filter block, cuvette, tubing)
1.21 Warning: Last calibration of sample pump towards analysis date was more than 90 days. (01-Jan-0001)
1.21.1 Solution - Check if the pumped analysis volume is the real volume by weighing
1.23 Warning: FWS-L and -R might indicate an air bubble problem 38% (100%=perfect)
1.23.1 Solution - Pinch valve to get rid of the air bubble or blockage by something different
1.27 Warning: SWS HS : Nr of particles with signal height <10 mV is high (92 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.27 Warning: FL Yellow HS : Nr of particles with signal height <10 mV is high (90 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.27 Warning: FL Orange HS : Nr of particles with signal height <10 mV is high (75 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
Disclaimer
Any use of the provided data and results or information provided by this website is on full own risk and responsibility. Those who gave access to this website or those who provided the results represented on the website cannot be held responsible for any damage or consequences that could arise from the use of this information. This LIVE system is empowered by the organization in the tab, CytoBuoy BV and Thomas Rutten Projects. TRP Copyright 2022