Online automatic phytoplankton monitoring at several locations performed by flowcytometry (CytoSense-CytoSub)
< HOME  TO OVERVIEW  1 TO OVERVIEW  2 >

Live Results Endeavour
To overview 2
nano_cend15_19 2020-02-07 22u03.cyz

----- Processed at : 26-Sep-2020 20:11:20 -----

1. The cytometric analysis is performed on machine Cefas
2. The CytoUSB software used is CytoUSB Version: 5.7.1.0
3. The filename is nano_cend15_19 2020-02-07 22u03.cyz
4. The analysis is performed on 07-02-2020 time 22:03:29
5. The number of measured particles is 67450
6. The total concentration found (e.g. only SWS trigg.) is 22 particles/microliter
7. The concentration found (all used triggers) is 22 particles/microliter
8. The preconcentration (before analysis) found (all used triggers) is 20 particles/microliter
9. The flush time found (between samples) is 180 seconds
10. The effective analysis volume is 3081.1958 microliter
11. The pumped volume is 5178 microliter
12. The particle-rate is 189 particles passing the laser beam/second
13. The used sample flow rate is 8.6324 microliter/second
14. The used block size memory is 4 kB
15. Trigger 1: FL Red at threshold level 21 mV
16. Trigger 2: not used at threshold level -999 mV
17. Temperature sensor PMT is : -999 C
18. Temperature sensor Sheath is : -999 C
19. Temperature sensor Sheath is : 22 C
20. Pressure sensor difference is : -999 mBar
21. Pressure sensor absolute is : -999 mBar
22. Sample pump 1: Last calibration of the sample pump was : 01-Jan-0001
23. Alignment check 1 by curvature detector is 1.4816 (1=perfect)
24. Alignment check 2 by curvature detector is 18% (100%=perfect)
25. Saturated signals (flattened peaks) occur when the dynamic range of the (HS)detector is reached
26. FWS: Nr of saturated signals is 0 (0% of 67450 measured particles)
27. FWS: Maximum signal height found is 4573 mV
28. FWS: Nr of particles with signal height <10 mV is high (14105 part. (21 %))
29. SWS HS: Nr of saturated signals is 0 (0% of 67450 measured particles)
30. SWS HS: Maximum signal height found is 4137 mV
31. SWS HS: Nr of particles with signal height <10 mV is high (56074 part. (83 %))
32. FL Yellow HS: Nr of saturated signals is 0 (0% of 67450 measured particles)
33. FL Yellow HS: Maximum signal height found is 2199 mV
34. FL Yellow HS: Nr of particles with signal height <10 mV is high (62816 part. (93 %))
35. FL Orange HS: Nr of saturated signals is 0 (0% of 67450 measured particles)
36. FL Orange HS: Maximum signal height found is 4026 mV
37. FL Orange HS: Nr of particles with signal height <10 mV is high (49083 part. (73 %))
38. FL Red HS: Nr of saturated signals is 0 (0% of 67450 measured particles)
39. FL Red HS: Maximum signal height found is 4289 mV
40. FL Red HS: Nr of particles with signal height <10 mV is high (25 part. (0 %))
41. FWS: Nr of block shaped signals is 196 (0.29059%)
42. FL Red HS: Nr of block shaped signals is 63 (0.093403%)
43. SWS HS: Nr of block shaped signals is 41 (0.060786%)
44. Suspect nr: 0=low, 10=high
45. Maximum FWS: suspect = 0
46. Maximum SWS HS: suspect = 0
47. Maximum FL Yellow HS: suspect = 1
48. Maximum FL Orange HS: suspect = 1
49. Maximum FL Red HS: suspect = 0
50. SWS HS: PMT level is 0
51. FL Yellow HS: PMT level is 0
52. FL Orange HS: PMT level is 0
53. FL Red HS: PMT level is 0

-------------------------- Warnings info --------------------------
1.05 Warning: Your effective analysis volume is low compared to the pumped analysis volume
1.05.1 Solution - Set blocksize memory in CytoUSB higher, used=4k (default=4k) e.g. to 8k
1.09 Warning: This (default) low block size of 4kB in CytoUSB will decrease the effective analysis volume
1.09.1 Solution - Increase block size memory if the total particle rate is high enough (but prevent time outs)
1.15 Warning: Pressure difference is high.
1.15.1 Solution - Check if there is something blocking from sample inlet to sheath (filter block, tubing)
1.16 Warning: Inner absolute pressure in fluidics towards outside is high.
1.16.1 Solution - Check if there is something blocking in the fluidics pathways (filter block, cuvette, tubing)
1.17 Warning: Last calibration of sample pump towards analysis date was more than 90 days. (01-Jan-0001)
1.17.1 Solution - Check if the pumped analysis volume is the real volume by weighing
1.18 Warning: FWS-L and -R detector signals are quite different (ratio 1.4816(1=perfect)) and might indicate an alignment problem
1.18.1 Solution - Probably you have an air bubble in your system. Pinch valve to get rid of it
1.18.2 Solution - Improve alignment of FWS-L and -R detector towards sample core
1.19 Warning: FWS-L and -R might indicate an air bubble problem 18% (100%=perfect)
1.19.1 Solution - Pinch valve to get rid of the air bubble or blockage by something different
1.23 Warning: SWS HS : Nr of particles with signal height <10 mV is high (83 %)
1.23.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.23.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.23.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.23 Warning: FL Yellow HS : Nr of particles with signal height <10 mV is high (93 %)
1.23.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.23.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.23.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.23 Warning: FL Orange HS : Nr of particles with signal height <10 mV is high (73 %)
1.23.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.23.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.23.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
Disclaimer
Any use of the provided data and results or information provided by this website is on full own risk and responsibility. Those who gave access to this website or those who provided the results represented on the website cannot be held responsible for any damage or consequences that could arise from the use of this information. Powered by TRP © 2016