Live Results Eijsden
To overview 2
EasyClus check & Info of last file

----- Processed at : 23-Apr-2024 18:30:06 -----

1. The cytometric analysis is performed on machine Lobith
2. The CytoBuoy instrument serial-nr is CS-2021-106
3. The CytoUSB software used is CytoUSB Version: 5.11.3.1
4. LASER : OBIS LS 488-60 : Used at 50 mW
5. LASER : OBIS LS 552-60 : Used at 50 mW
6. The filename is EIJ2_SWS15_FLRed10_auto_uls_IMG_600s 2024-04-23 18h14.cyz
7. The analysis is performed on 23-04-2024 time 18:17:15
8. The number of measured particles is 20115
9. The total concentration found (e.g. only SWS trigg.) is 602 particles/microliter
10. The concentration found (all used triggers) is 11 particles/microliter
11. The preconcentration (before analysis) found (all used triggers) is 11 particles/microliter
12. The flush time found (between samples) is 130 seconds
13. The effective analysis volume is 1836.3378 microliter
14. The pumped volume is 1848 microliter
15. The particle-rate is 1854 particles passing the laser beam/second
16. The used sample flow rate is 3.08 microliter/second
17. Trigger 1: SWS HS at threshold level 15 mV
18. Trigger 2: Maximum FL Red HS at threshold level 10
19. Temperature sensor PMT is : 28.4 C
20. Temperature sensor Sheath is : 23.1 C
21. Temperature sensor Laser is : 28.4 C
22. Temperature sensor System is : 24.8 C
23. Pressure sensor difference is : -601 mBar
24. Pressure sensor absolute is : 457 mBar
25. Sample pump 1: Last calibration of the sample pump was : 01-Jan-1990
26. Alignment check 1 by curvature detector is 0.91024 (1=perfect)
27. Alignment check 2 by curvature detector is 64.2% (>70%=perfect)
28. Saturated signals (flattened peaks) occur when the dynamic range of the (HS)detector is reached
29. FWS: Nr of saturated signals is 1 (0% of 20115 measured particles)
30. FWS: Maximum signal height found is 14464 mV
31. FWS: Nr of particles with signal height <10 mV is high (298 part. (1 %))
32. SWS HS: Nr of saturated signals is 1 (0% of 20115 measured particles)
33. SWS HS: Maximum signal height found is 7360 mV
34. SWS HS: Nr of particles with signal height <10 mV is high (0 part. (0 %))
35. FL Yellow HS: Nr of saturated signals is 0 (0% of 20115 measured particles)
36. FL Yellow HS: Maximum signal height found is 2752 mV
37. FL Yellow HS: Nr of particles with signal height <10 mV is high (12204 part. (61 %))
38. FL Orange HS: Nr of saturated signals is 1 (0% of 20115 measured particles)
39. FL Orange HS: Maximum signal height found is 7360 mV
40. FL Orange HS: Nr of particles with signal height <10 mV is high (3198 part. (16 %))
41. FL Red HS: Nr of saturated signals is 20 (0% of 20115 measured particles)
42. FL Red HS: Maximum signal height found is 7360 mV
43. FL Red HS: Nr of particles with signal height <10 mV is high (514 part. (3 %))
44. FL Orange HS: Nr of block shaped signals is 203 (1.0092%)
45. FL Red HS: Nr of block shaped signals is 62 (0.30823%)
46. SWS HS: Nr of block shaped signals is 23 (0.11434%)
47. FWS: Nr of block shaped signals is 21 (0.1044%)
48. FL Yellow HS: Nr of block shaped signals is 15 (0.074571%)
49. Suspect nr: 0=low, 10=high
50. Maximum FWS: suspect = 0
51. Maximum SWS HS: suspect = 0
52. Maximum SWS LS: suspect = 3
53. Maximum FL Yellow HS: suspect = 3
54. Maximum FL Yellow LS: suspect = 0
55. Maximum FL Orange HS: suspect = 3
56. Maximum FL Red HS: suspect = 0
57. Maximum FL Red LS: suspect = 0
58. SWS HS: PMT level is 49
59. SWS LS: PMT level is 48
60. FL Yellow HS: PMT level is 94
61. FL Yellow LS: PMT level is 76
62. Fl Orange: PMT level is 105
63. FL Red HS: PMT level is 110
64. FL Red LS: PMT level is 81
65. IIF: Imaging in Flow pictures available. Nr of images is 1000

-------------------------- Warnings info --------------------------
1.21 Warning: Last calibration of sample pump towards analysis date was more than 90 days. (01-Jan-1990)
1.21.1 Solution - Check if the pumped analysis volume is the real volume by weighing
1.24 Warning: SWS LS: Maximum is 576 mV if HS is around 4500 mV and too high.
1.24.1 Solution - Put LS PMT lower. Maximum LS around 50 mV (if HS is around 4500 mV).
1.24 Warning: FL Yellow LS: HS-LS ratio is 22 and too low.
1.24.1 Solution - Put LS PMT lower. Ratio HS/LS should be around 100.
1.24 Warning: FL Yellow LS: Maximum is only 7 mV (if HS is around 4500 mV) and too low.
1.24.1 Solution - Put LS PMT higher. Maximum LS around 50 mV (if HS is around 4500 mV).
1.24 Warning: FL Yellow LS: HS-LS ratio is 32 and too low.
1.24.1 Solution - Put LS PMT lower. Ratio HS/LS should be around 100.
1.24 Warning: FL Red LS: Maximum is 118 mV if HS is around 4500 mV and too high.
1.24.1 Solution - Put LS PMT lower. Maximum LS around 50 mV (if HS is around 4500 mV).
1.24 Warning: FL Red LS: HS-LS ratio is 62 and too low.
1.24.1 Solution - Put LS PMT lower. Ratio HS/LS should be around 100.
1.27 Warning: FL Yellow HS : Nr of particles with signal height <10 mV is high (61 %)
1.27.1 Solution - Increase, set higher trigger threshold (to prevent measuring noise)
1.27.2 Solution - Decrease, set lower particle sample rate. Maybe caused by an overload of particles/ noise in your sample
1.27.3 Solution - Clean sheath, clean cuvette, clean tubing? (noise can be introduced by dirty walls)
1.25 Warning: Suspected particles data blocks found!
1.29.1: Maximum SWS LS: suspect = 3
1.29.11 Solution - Particle rate too high?
1.29.21 Solution - Trigger too sensitive?
1.29.31 Solution - Details look at 'DATA DIAGNOSIS' in 'Check CytoSense-Diagnosis-Meta Data'
1.29.2: Maximum FL Yellow HS: suspect = 3
1.29.12 Solution - Particle rate too high?
1.29.22 Solution - Trigger too sensitive?
1.29.32 Solution - Details look at 'DATA DIAGNOSIS' in 'Check CytoSense-Diagnosis-Meta Data'
1.29.3: Maximum FL Orange HS: suspect = 3
1.29.13 Solution - Particle rate too high?
1.29.23 Solution - Trigger too sensitive?
1.29.33 Solution - Details look at 'DATA DIAGNOSIS' in 'Check CytoSense-Diagnosis-Meta Data'
Disclaimer
Any use of the provided data and results or information provided by this website is on full own risk and responsibility. Those who gave access to this website or those who provided the results represented on the website cannot be held responsible for any damage or consequences that could arise from the use of this information. This LIVE system is empowered by Rijkswaterstaat Laboratorium, CytoBuoy BV and Thomas Rutten Projects. TRP © 2024